Procom VOR/ILS Analyzer EVS200 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Sprzęt komputerowy Procom VOR/ILS Analyzer EVS200. Procom VOR/ILS Analyzer EVS200 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - VOR/ILS Analyzer EVS200

0798.1988.12-08User ManualVOR/ILS Analyzer EVS2000796.1800.02

Strona 2

VOR/ILS Analyzer EVS2000796.1800.02 E-8 1Chapter 1: Operator InformationImportant operator informationThe symbols used in this description have the fo

Strona 3

VOR/ILS Analyzer EVS2000796.1800.02 E-8 2Chapter 1: Operator InformationUnpacking1. Unpack the VOR/ILS Analyzer EVS200.2. Remove the protective covers

Strona 4

VOR/ILS Analyzer EVS2000796.1800.02 E-8 3Chapter 1: Operator InformationApplication of the unitThe VOR/ILS Analyzer EVS200 is used for checking terres

Strona 5 - Safety Instructions

VOR/ILS Analyzer EVS2000796.1800.02 E-8 4Chapter 1: Operator InformationCharacteristics of the unitRemarkable characteristics of the VOR/ILS Analyzer

Strona 6

5E-8Chapter 1: Operator InformationUnit layout frontviewBNC socketinput level:VSWR:max. +15 dBm / 50 Ω<1.5antenna inputturning direction for increa

Strona 7 - 3. Operation

6E-8Chapter 1: Operator InformationUnit layout rearviewexternal VDC inputWhen an external VDC supply is connectedthe installed battery (option) is swi

Strona 8 - 4. Interfaces

VOR/ILS Analyzer EVS2000796.1800.02 E-8 7Chapter 2: Preparation for OperationPreparation for operationPlacing the unitThe VOR/ILS Analyzer EVS200 can

Strona 9 - 6. Technical Specification

VOR/ILS Analyzer EVS2000796.1800.02 E-8 8Chapter 2: Preparation for OperationDC connectionOnly connect the unit when the minus pole of the batteryis c

Strona 10

VOR/ILS Analyzer EVS2000796.1800.02 E-8 9Chapter 2: Preparation for OperationConnection of Signal/Control Inputs/OutputsAntenna connectionVia the RF I

Strona 11 - Unpacking

VOR/ILS Analyzer EVS2000796.1800.02 E-8 10Chapter 2: Preparation for OperationExternal AF inputVia the AF input (AF-EXT) at the rear side the unit can

Strona 12 - Application of the unit

0798.1988.12-08VOR/ILS Analyzer EVS200Issue: 01.2002Version: E8Copying of this document as well as any other utilization and communication of its

Strona 13 - Characteristics of the unit

VOR/ILS Analyzer EVS2000796.1800.02 E-8 11Chapter 3: OperationOperationSwitch On/switch Off the unit at mains supply With the mains switch at the rea

Strona 14 - Unit layout frontview

VOR/ILS Analyzer EVS2000796.1800.02 E-8 12Chapter 3: OperationSelftest (BITE)The selftest checks: all operating voltages, host processor, memory,

Strona 15 - Unit layout rearview

VOR/ILS Analyzer EVS2000796.1800.02 E-8 13Chapter 3: OperationOperation of the unitBecause of the many universal measuring functions of the VOR/ILS An

Strona 16 - Preparation for operation

VOR/ILS Analyzer EVS2000796.1800.02 E-8 14Chapter 3: OperationSETUP-ModeIn the following table all possible parameters are listed and its functions ar

Strona 17 - DC connection

VOR/ILS Analyzer EVS2000796.1800.02 E-8 15Chapter 3: OperationOperating instructions of SETUP-Mode

Strona 18 - =oscilloscope

VOR/ILS Analyzer EVS2000796.1800.02 E-8 16Chapter 3: OperationILS-ModeOperating instructions of ILS-Mode

Strona 19 - =external receiver

VOR/ILS Analyzer EVS2000796.1800.02 E-8 17Chapter 3: OperationOperating instructions of Y /t setup menu

Strona 20 - Operation

VOR/ILS Analyzer EVS2000796.1800.02 E-8 18Chapter 3: OperationOperating instructions of STORE DDM menu

Strona 21 -  DSP processor

VOR/ILS Analyzer EVS2000796.1800.02 E-8 19Chapter 3: Operation

Strona 22 - Operation of the unit

VOR/ILS Analyzer EVS2000796.1800.02 E-8 20Chapter 3: OperationSignal parameters on ILS displayFollowing signal parameters are indicated on the display

Strona 23 - SETUP-Mode

EC Certificate of ConformityCertificate No.: 0201This is to certify thatEquipment type: EVS200Designation: VOR/ILS AnalyzerOrder No.: 796.1800.02compl

Strona 24

VOR/ILS Analyzer EVS2000796.1800.02 E-8 21Chapter 3: OperationVOR-ModeOperating instructions of VOR-Mode

Strona 25 - ILS-Mode

VOR/ILS Analyzer EVS2000796.1800.02 E-8 22Chapter 3: OperationSignal parameters on VOR displayFollowing signal parameters are indicated on the display

Strona 26

VOR/ILS Analyzer EVS2000796.1800.02 E-8 23Chapter 3: OperationBEACON-ModeOperating instructions of BEACON-Mode

Strona 27

VOR/ILS Analyzer EVS2000796.1800.02 E-8 24Chapter 3: OperationSignal parameters on BEACON displayFollowing signal parameters are indicated on the disp

Strona 28

VOR/ILS Analyzer EVS2000796.1800.02 E-8 25Chapter 3: Operation∆∆∆∆ Level-ModeOperating instructions of ∆∆∆∆ Level-Mode

Strona 29

VOR/ILS Analyzer EVS2000796.1800.02 E-8 26Chapter 3: OperationStoring and recalling reference levelsTo occupy storage capacaty or to call up memory th

Strona 30 - VOR-Mode

VOR/ILS Analyzer EVS2000796.1800.02 E-8 27Chapter 3: Operation

Strona 31

VOR/ILS Analyzer EVS2000796.1800.02 E-8 28Chapter 3: OperationSignal parameters on ∆∆∆∆ Level DisplayFollowing signal parameters are indicated on the

Strona 32 - BEACON-Mode

VOR/ILS Analyzer EVS2000796.1800.02 E-8 29Chapter 3: OperationSpectrum-ModeOperating instructions of Spectrum-ModeThe spectrum analysis of the VOR/ILS

Strona 33

VOR/ILS Analyzer EVS2000796.1800.02 E-8 30Chapter 3: OperationRS-232-Interface operationAll important unit functions can be remote controlled via the

Strona 35

VOR/ILS Analyzer EVS2000796.1800.02 E-8 31Chapter 3: OperationUnit related control commandsCommand RF (frequency in kHz)command functionRF (frequency

Strona 36

VOR/ILS Analyzer EVS2000796.1800.02 E-8 32Chapter 3: OperationCommand EC1command functionEC1 <echoon> selection of communicationEVS200 retransmi

Strona 37 - ∆ Level Display

VOR/ILS Analyzer EVS2000796.1800.02 E-8 33Chapter 3: OperationCommand REScommand functionRES <RESET> Master-RESETWith the command RES a unit mas

Strona 38 - Spectrum-Mode

VOR/ILS Analyzer EVS2000796.1800.02 E-8 34Chapter 3: OperationCommand VERcommand functionVER <Version> query of version number and date of theEV

Strona 39 - RS-232-Interface operation

VOR/ILS Analyzer EVS2000796.1800.02 E-8 35Chapter 3: OperationCommand MIcommand functionMI <Mode-ILS> ILS mode settingThe command MI enables the

Strona 40 - Unit related control commands

VOR/ILS Analyzer EVS2000796.1800.02 E-8 36Chapter 3: OperationCommand DD1command functionDD1 <DDM information>query of the DDM value in µµµµAThe

Strona 41

VOR/ILS Analyzer EVS2000796.1800.02 E-8 37Chapter 3: OperationCommand AM2command functionAM2 <AM modulation depth (90 Hz)> query of AM modulatio

Strona 42

VOR/ILS Analyzer EVS2000796.1800.02 E-8 38Chapter 3: OperationCommand BW0command functionBW0 <bandwidth information > 8-kHz bandwidth settingThe

Strona 43

VOR/ILS Analyzer EVS2000796.1800.02 E-8 39Chapter 3: OperationCommand FA1command functionFA1 <Fast DDM measuring on> Activating the Fast DDM mea

Strona 44 - Mode related control commands

VOR/ILS Analyzer EVS2000796.1800.02 E-8 40Chapter 3: OperationCommand FA2command functionFA2 <Fast DDM measuring on> Activating the Fast DDM mea

Strona 45

Safety Instructions095.1000 Sheet 17This unit has been designed and tested in accordance with the EC Certificate of Conformity and has left themanufac

Strona 46

VOR/ILS Analyzer EVS2000796.1800.02 E-8 41Chapter 3: OperationCommand FA3command functionFA3 <Fast DDM measuring on> Activating the Fast DDM mea

Strona 47 - Command BW1

VOR/ILS Analyzer EVS2000796.1800.02 E-8 42Chapter 3: OperationCommand FA0command functionFA0 <fast DDM measuring off> deactivating the fast DDM

Strona 48 - Command FA1

VOR/ILS Analyzer EVS2000796.1800.02 E-8 43Chapter 3: OperationCommand AF8command functionAF8 <AF-frequency ID> AF-frequency ID queryThe command

Strona 49 - Command FA2

VOR/ILS Analyzer EVS2000796.1800.02 E-8 44Chapter 3: OperationVOR-ModeCommand BEcommand funcionBE <Bearing Angle> bearing angle queryThe command

Strona 50

VOR/ILS Analyzer EVS2000796.1800.02 E-8 45Chapter 3: OperationCommand FM0command functionFM0 <FM-deviation)> FM-deviation queryThe command FM0 r

Strona 51

VOR/ILS Analyzer EVS2000796.1800.02 E-8 46Chapter 3: OperationCommand D2command functionD2 <Display-information> switching over to Display2The c

Strona 52

VOR/ILS Analyzer EVS2000796.1800.02 E-8 47Chapter 3: OperationCommand AM9command functionAM9 <AM-modulation depth (voice)> query AM-modulation d

Strona 53

VOR/ILS Analyzer EVS2000796.1800.02 E-8 48Chapter 3: OperationCommand FA3command functionFA3 <Fast measuring on> Activating the Fast measurement

Strona 54

VOR/ILS Analyzer EVS2000796.1800.02 E-8 49Chapter 3: Operation∆∆∆∆ Level-ModeCommand SRcommand functionSR <Set REF> reference level settingThe c

Strona 55

VOR/ILS Analyzer EVS2000796.1800.02 E-8 50Chapter 3: OperationBeacon-ModeCommand AM4command functionAM4 <AM-modulation depth (F1-range)> query o

Strona 56

Safety Instructions095.1000 Sheet 18Ensure that the connections with information tech-nology equipment comply with IEC950/EN60950.11. Lithium batterie

Strona 57

VOR/ILS Analyzer EVS2000796.1800.02 E-8 51Chapter 3: OperationCommand AM7Kommando FunktionAM6 <AM-modulation depth ID> query of AM-modulation de

Strona 58

VOR/ILS Analyzer EVS2000796.1800.02 E-8 52Chapter 3: OperationCommand AF6command functionAF6 <AF-frequency range F3> query of AF-frequency 400 H

Strona 59 - Beacon-Mode

VOR/ILS Analyzer EVS2000796.1800.02 E-8 53Chapter 4: InterfacesInterfacesAntenna inputlabeling figure descriptionRF INPUT 50 Ω input level: 0 dBm / 50

Strona 60

VOR/ILS Analyzer EVS2000796.1800.02 E-8 54Chapter 4: InterfacesExternal AF inputlabeling figure descriptionAF EXT.input level: 1 to 2 Vrms/50Ωfrequenc

Strona 61 - Command AF7

VOR/ILS Analyzer EVS2000796.1800.02 E-8 55Chapter 5: ServiceServiceTo ensure a unit’s repair as quick as possible a defect VOR/ILS Analyzer EVS200must

Strona 62 - Interfaces

VOR/ILS Analyzer EVS2000796.1800.02 E-8 56Chapter 6: Technical SpecificationTechnical dataReceiver sectionFrequency range ...

Strona 63

VOR/ILS Analyzer EVS2000796.1800.02 E-8 57Chapter 6: Technical SpecificationGlideslope modeMeasurement error at30 to 50% modulation...

Strona 64 - Warranty

VOR/ILS Analyzer EVS2000796.1800.02 E-8 58Chapter 6: Technical SpecificationGeneral dataPower supply ...100 to 240 VAC(wi

Strona 65 - Technical data

VOR/ILS Analyzer EVS2000796.1800.02 E-8 59Chapter 6: Technical SpecificationAccessoriesDescription Ident No.Bag for EVS200 TEVS 0798-4264Alu-tripod fo

Strona 66 - VOR signal analysis

VOR/ILS Analyzer EVS2000796.1800.02 E-8 0.1ContentsContentsChapter Page1. Operator Information- Important operator information 1- Unpacking 2- Applica

Strona 67 - General data

VOR/ILS Analyzer EVS2000796.1800.02 E-8 0.2ContentsChapter Page- ILS-Mode 16- Operating instructions of ILS-Mode 16- Operating instructions of Y / t s

Strona 68 - Description Ident No

VOR/ILS Analyzer EVS2000796.1800.02 E-8 0.3ContentsChapter Page5. Service- Service 55- Warranty 556. Technical Specification- Technical data 56- Gener